Instrument
Calibration
Main Menu
Instr. Calibration?
Instrument Cal. Menu
Perform Instr. Cal.?
Calibrating the PCS:
To access the Instrument Calibration Menu, press MENU
SCROLL until the Instrument Calibration prompt is displayed
(Figure 20).
Press YES at the prompt.
The Instrument Calibration Menu will open.
Press YES at the Perform Instrument Calibration prompt
(Figure 21).
Important Notes:
Minimally, the Instrument Calibration should be performed every 30 days or prior to the next use, whichever is
later.
The Instrument is calibrated using the PCS Instrument Calibrator Kit, which contains 4 vials with calibrator
values: CAL A, B, C and D.
The contents of the kit are subject to degradation with time, temperature and exposure to light.
Figure 20
Figure 21
PCS Quickstart-15A3179_PCS Quickstart 15A3179 Rev 4/3/13 2:02 PM Page 16