Instrument
Calibration
Main Menu
Instr. Calibration?
Instrument Cal. Menu
Perform Instr. Cal.?
Calibrating the PCS:
• To access the Instrument Calibration Menu, press MENU
SCROLL until the Instrument Calibration prompt is displayed
(Figure 20).
• Press YES at the prompt.
• The Instrument Calibration Menu will open.
• Press YES at the Perform Instrument Calibration prompt
(Figure 21).
Important Notes:
• Minimally, the Instrument Calibration should be performed every 30 days or prior to the next use, whichever is
later.
• The Instrument is calibrated using the PCS Instrument Calibrator Kit, which contains 4 vials with calibrator
values: CAL A, B, C and D.
• The contents of the kit are subject to degradation with time, temperature and exposure to light.
Figure 20
Figure 21