EasyManua.ls Logo

Artel PCS - Managing Calibration Runs

Artel PCS
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Always refer to the PCS Procedure Guide for further instruction
Another Cal w/Vial?
(Yes/No)
Use New Vial?
(Yes/No)
End of Vial
The number of data points that can be collected in each Blank
vial is dependent on the sample volume and will range from 1
point to 40.
Each time a pipette calibration run is completed, the
Instrument will display the remaining number of samples of the
selected volume that may be dispensed into the vial.
In any one Blank vial, calibrations may be performed at the
same volume, at different volumes with the same Sample
Solution or at different volumes with different Sample
Solutions.
To use the same vial after the completion of the first
calibration, press YES at the first prompt (Figure 17). The
calibration cycle will restart.
When the maximum capacity of the vial has been reached,
alternating prompts will flash (Figures 18 & 19).
If more data points are required, answer YES to the Use New
Vial prompt. The Instrument will prompt to remove the old vial
and insert a new one. This process may be repeated if
needed.
Figure 17
Figure 18
Figure 19
Completing
Calibration
PCS Quickstart-15A3179_PCS Quickstart 15A3179 Rev 4/3/13 2:02 PM Page 15

Related product manuals