45 Wrong product combination:
Mixup between non compatible device and hand held
programming unit
-> Use the correct hand held programming unit (Pos. 1.31)
46 Handset error internal communication:
Invalid interchip communication inside the hand held
programming unit
-> Replace hand held programming unit (Pos. 1.31)
47 Internal communication error motor control:
Internal communication error motor control..
-> Replace motor control (Pos. 1.35)
48 User stoped the special function
49 Unknown error in the motor control:
-> Replace motor control (Pos. 1.35)
The threaded spindle and spindle nut is not
interchangeable with
ARTROMOT®-K4
serial number < 3000
.
Do not loosen the knurled handles completely for any
adjustment. For operation or transport, make sure that
the knurled handles are tight.
The frame is unstable:
Possible cause: Bolt / screws missing or loose. Tighten
the screws / bolts.
3.3 Others
Do not clean the casing or the support with grease
or oil.
Do not utilize any solvent for cleaning the
ARTROMOT®-K4.