(G3) Smart If the smart picker is switched on, the highest last temperature value will always be displayed and
stored. This feature is particularly useful when object temperature is not uniform across its dimension or the
pyrometer is not constantly viewing an object to be measured.
Smart Picker Functions
Smart picker can be turn ON & OFF by using the software. When Peak picker is ON, the peak picker menu is enabled
for setting of the parameters like decay rate function, reset below temperature and peak picker delay.
Accurate Sensors Technologies
14
(I)Decay rate The Decay rate range is 0.00 to 166.66°C/sec. or 0.00
to 300°F/sec. depending upon °F/°C units selection. The slowest
Decay rate is 0 degrees per sec. This feature helps to eliminate
erratic measurements and allows the peaked value to decay down
to lower process temperature values as they occur. Decay rate is set
to retain peak measured temperature value and ignore momentary
decreases in measured temperature.
(ll)Delay Time This function set the delay time in sec. before peak
picker function starts. The delay time is selectable in the range 0.02
to 10.00 sec. Zero (0) turns delay time OFF. This function is used to
delay the start of peaking action for upto 10 sec. following the
detection of leading edge of a new target.
Temp
Time
Delay Time (sec.)
Temp
Time
Without RBT
RBT with Decay Rate
Temp
Time
RBT
Outer
with Peak Picker
Current Measuring
Value
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate
Outer
with Peak Picker
Current Measuring
Value