EasyManua.ls Logo

Astralpool Smart+ 60 - Calibration of the pH sensor

Astralpool Smart+ 60
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
Fig. 30
6.2.2. “STANDARD MODE
“STANDARD” MODE allows the precise calibration of the pH-sensor using two calibration solutions of pH 7.0 and 4.0, however this method
requires that the pH-sensor is removed from the installation.
PROCEDURE:
IMPORTANT: before closing the by-pass valves, stop system from control panel (see Section 5.1).
1. Extract the pH-sensor from the holder and wash it with tap water.
Fig. 31
2. Press the “CAL” [18] key until the equipment beeps. Without releasing the key, press “SET” [21] for a few seconds until the pH
screen displays a blinking “7.0”. The calibration light “7.0” [17] will remain on.
3. Gently shake the sensor to remove any possible drops of water and insert in the standard pH=7.0 solution (green). Gently shake
for a few seconds and press “CAL” [18]. Once stabilised, the screen [15] will display a blinking “4.0”. The calibration light “4.0” will
remain on [17].
Fig. 32

Related product manuals