DC Voltage (Auto ranging)
Range Resolution Accuracy
600mV 0.1mV ±(0.8% of rdg + 5dgts)
6V 1mV ±(0.8% of rdg + 3dgts)
60V 10mV
600V 100mV
1000V 1V ±(1.0% of rdg + 5dgts)
Input Impedance: 10MΩ
Overload Protection: 600V DC/AC RMS
Max. Input voltage: 600V DC
Measure DCV 1000V under CATII
AC Voltage (Auto ranging)
Range Resolution Accuracy
600mV 0.1mV ±(1.2% of rdg + 8dgts)
6V 1mV
±(1.2% of rdg + 6dgts) 60V 10mV
600V 100mV
750V 1V ±(1.2% of rdg + 8dgts)
Input Impedance: 10MΩ
Frequency Range: 40Hz ~ 400Hz
Overload Protection: 600V DC/AC rms
Measure AC V 750V under CATII CAT III
Max. Input voltage: 600V AC RMS CATIII