Guidance and manufacturer´s declaration electromagnetic emission
2
The A310 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
customer or the user of A310 Pulse Oximeter should assure that it is used in such an environment.
3
Emissions test
Compliance
Electromagnetic environment - guidance
4
RF emissions
CISPR 11
Group 1
The A310 Pulse Oximeter uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
5
RF emissions
CISPR 11
Class B
The A310 Pulse Oximeter is suitable for use in
all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supplynetwor
k that supplies buildings used for domesticpurposes.
6
Harmonic emissions
IEC 61000-3-2
N/A
7
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
N/A
Guidance and manufacturer's declaration electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration electromagnetic immunity
TheA310 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
customer or the user of the A310 Pulse Oximeter should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
N/A
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common mode
N/A
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
0 % UT; 0,5 cycle g)
At 0°, 45°, 90°, 135°, 180°,
N/A
Mains power quality should be that of a
typical commercial or hospital