12
/
15
for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
The A310 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
customer or the user of A310 Pulse Oximeter should assure that it is used in such an environment.
Electromagnetic environment - guidance
The A310 Pulse Oximeter uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
The A310 Pulse Oximeter is suitable for use in
all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supplynetwor
k that supplies buildings used for domesticpurposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
TheA310 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
customer or the user of the A310 Pulse Oximeter should assure that it is used in such an environment.
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
± 1 kV differential
mode
± 2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
0 % UT; 0,5 cycle g)
At 0°, 45°, 90°, 135°, 180°,
Mains power quality should be that of a
typical commercial or hospital