EasyManua.ls Logo

Atten ST-862D - Switching Temperature Unit; Temperature Unit Selection; Temperature Calibration; Calibration Procedure

Atten ST-862D
27 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Temperature Calibration
0 5
C
C
A
L
SET
- 0 5
C
C
A
L
SET
Under menu CAL mode, press “▲”or“▼”key to enable temperature calibration
function, whose calibration range is -50℃~ 50℃(-90℉~90℉). Calibration value will be
negative as actual-measured temperature is lower than setting temperature; Calibration
value will be positive as actual-measured temperature is higher than setting temperature.
FIG. 1-12 FIG. 1-13
6. Button Beeping Function
Under menu BL mode, user can turn on/ off the beeping function by pressing “▲”or
“▼”key.Refer Pic.1-14, OFF disable beeping function;Refer Pic.1-15, ON enables beeping
function.
OF F
REAL SET
0
REAL SET
FIG. 1-14 FIG. 1-15
4. Switching Temperature Unit
- C-
C
SET
C
-
F
- F -
F
SET
C
-
F
FIG. 1-10 FIG. 1-11
7
Note: User need calibrate temperature by adjusting its value if the displayed temperature is much different with
actual-measured temperature after changing a new heater or tips, its calibration method be refered as below:
1.Set this handle required calibration as a suitable temperature like 350/ 662;
2.After this setting temperature is stable, use temperature tester to measure the actual temperature of its tip,
for example, its actual-measured temperature is 365/ 689;
3.We can get conclusion that the actual-measured temperature compar
ed to the setting temperature is
15/ 27℉ higher;
4.Then press “” key to calibrate temperature.

Related product manuals