14
.
Technical Data
14
.
1
6150AD
Dose Rate Meter 6150...
AD1
AD3
AD5
AD1/H, AD1/E
AD3/H, AD3/E
AD5/H, AD5/E
AD2
AD4
AD6
AD2/H, AD2/E
AD4/H, AD4/E
AD6/H, AD6/E
Detector
gamma tube ZP1310 or
equivalent, energy
compensated, effective length
16 mm, sensitivity at Cs-137
approx. 500 pulses per µSv
gamma tube ZP1210 or
equivalent, energy
compensated, effective length
40 mm, sensitivity at Cs-137
approx. 5800 pulses per µSv
Measuring quantity:
photon dose (rate) equivalent H
X
or
ambient dose (rate) equivalent H*(10)
H
X
H*(10)
H
X
H*(10)
Energy dependence:
nominal energy range
45 keV to 3 MeV
60 keV to 1.3 MeV
deviation referred to Cs-137
max. ±30%
max. ±30%
Directional dependence:
nominal angular range
±45° around preferential direction
deviation referred to preferential
direction at the same energy
max. ±20%
max. ±20%
Energy and directional dependence,
deviation for all energies and
directions referred to Cs-137 at
preferential direction
max. ±40%
max. ±40%
Analog dose rate range
Digital dose rate range
1 µSv/h to 1000 mSv/h
0.0 µSv/h to 999 mSv/h
0.1 µSv/h to 10 mSv/h
0.00 µSv/h to 9.99 mSv/h
Instrumental background
< 20 nSv/h
< 20 nSv/h
Linearity of dose rate measurement
deviation max. ±10%, calibration with Cs-137
Overload: overrange will be indicated
up to:
50 Sv/h
500 mSv/h
Dose rate mean value:
digital range (see also sect. 6.2)
statistical accuracy
0.000 µSv/h to 9.999 mSv/h
digits flash as long as standard deviation is greater than 5%
Digital range for dose rate maximum
value (see also section 6.4)
6150AD1 (/H, /E):
0 µSv/h to 999 mSv/h
6150AD3/5 (/H, /E):
0.00 µSv/h to 999 mSv/h
6150AD2 (/H, /E):
0 µSv/h to 9.99 mSv/h
6150AD4/6 (/H, /E):
0.00 µSv/h to 9.99 mSv/h
Digital dose range (see also section
6.5)
6150AD1 (/H, /E):
0 µSv to 9.99 Sv
6150AD3/5 (/H, /E):
0.00 µSv to 9.99 Sv
6150AD2 (/H, /E):
0 µSv to 99.9 mSv
6150AD4/6 (/H, /E):
0.00 µSv to 99.9 mSv
03/2004
Operating Manual 6150AD
Page
43
Daimlerstrasse 27
Telefon
+49(0)6203-9503-00
http://www.automess.de
USt.-ID-Nr. / VAT ID No.
D-68526 Ladenburg
Telefax
+49(0)6203-9503-29
eMail: info@automess.de
DE 144 458 420

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Automess 6150AD1 and is the answer not in the manual?

Automess 6150AD1 Specifications

General IconGeneral
BrandAutomess
Model6150AD1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals