32
4 Automated Measurement Sequence - 9181
Na K Cl READY AUTOMATED
MEASUREMENT?
CHECK SS
VOLTAGE
Sample *
in Process
NO SAMPLE!
Na K Cl
151 5.1 114
NO
YES
No sample = 2No sample < 2
Probe moves to
upper position
Manual
Measurement
YES
Tray homes
Tray moves to
applicable cup
position
Probe moves
down to cup
position
Turntable
moves to mid
position
Probe moves to
lower position
End of Run
Home Tray
Na K Cl READY
Probe moves to
upper position
* applicable sample number