Questions and Answers

  • E
    Eric CummingsAug 11, 2025
    Why Awareness Technology Stat Fax 4200 Laboratory Equipment says 'Assay Cannot Be Edited!'?
    • S
      Shirley MedinaAug 12, 2025
      This message indicates that you are trying to edit an OEM preprogrammed test that has a flag set to prevent editing.
  • B
    butlerkimberlyAug 19, 2025
    Why does my Awareness Technology Stat Fax 4200 say '%Absorbance Mode should not be used with negative absorbances!'?
    • B
      Barry AndersonAug 19, 2025
      This message displays if you are in %Absorbance Mode and a standard is read which has a negative absorbance.
  • J
    Jennifer ChapmanAug 27, 2025
    What does '******' mean on my Awareness Technology Laboratory Equipment?
    • S
      snyderveronicaAug 27, 2025
      This indicates that a sample has exceeded the absorbance of the reference well in a log-logit test. Specifically, Curve Axes is Y=Logit(Abs), X=Log(Conc), and Absorbance is greater than the reference (zero) absorbance value. It will also appear if a curve with a log axis tries to log a zero, a negative number, or an invalid curve has been read. It prints in the Concentration column whenever an error occurs while calculating the concentration.
  • S
    Scott JoyceSep 3, 2025
    Why does Awareness Technology Laboratory Equipment say 'Do not select the same filter for primary and differential'?
    • W
      Willie GrossSep 3, 2025
      This message appears on Page 1 of Test Definition when the same wavelengths are entered for the Primary and the Differential filters.

Awareness Technology Stat Fax 4200 Specifications

General IconGeneral
TypeMicroplate Reader
Light SourceTungsten halogen lamp
Wavelength Range400 - 700 nm
Aspiration VolumeNot applicable
Power Requirements100-240 V, 50/60 Hz
DisplayLCD
PrinterOptional
Read ModeEndpoint and kinetic
Plate Type96-well microplates
InterfaceRS-232
Weight14 lbs (6.4 kg)