X. Error Codes
1. OUTZ - Zero point has drifted out of the acceptable
range. Try to perform a calibration adjustment. If the
problem persists, the load cell may be permanently
damaged due to overloading.
2. O-Ld - Overload. The maximum capacity has
been exceeded. Remove the extra weight to avoid
permanently damaging the load cell.
3. Lo - Batteries are low. Replace the batteries.
XI. Specications
AERO-100 AERO-650
Max. Capacity 100g 600g
Readability 0.01g 0.1g
Min. Weight 0.03g 0.2g
Power 2x1.5V“AAA”Alkaline
Dimensions 0.7 x 2.6 x 4.7 in.
Platform Dimensions 2.0 x 2.9 in.
Warranty 10 Year Limited Warranty
XII. Capacity/Readability Charts
AERO-100 AERO-650
100g x 0.01g
3.527oz x 0.001oz
3.215ozt x 0.001ozt
64.30dwt x 0.01dwt
500ct x 0.05ct
1543.2gn x 0.2gn
650g x 0.1g
22.92oz x 0.01oz
20.89ozt x 0.01ozt
417.95dwt x 0.1dwt
3250ct x 0.5ct
10031gn x 2gn