LB-Series Manual
4. To view the cummulative weight, press ON/MEM.
The total is displayed and memory cleared.
5. Press TARE/PRINT again to zero out the balance
and the scale is ready for normal weighing again.
IX. Calibration
Calibration and adjustment are performed at the factory.
However, the weighing range can shift slightly during
shipping. Also, due to changes in local acceleration of gravity
around the world, adjustment is recommended when moving
the scale from one location to another. To check calibration:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart below), adjustment is required.
Model Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
LB-501 0.03g 400g OIML M2
LB-1000 0.2g 400g OIML M2
LB-3000 0.2g 2000g OIML M2
X. Adjustment
1. Press and hold the MODE/CAL key until the display
shows “CAL”.
2. Press the MODE/CAL key again. The display will
ash “CAL” followed by the calibration weight value
(see chart).
3. Place the calibration weight on the center of the
weighing platform. Make sure you are using the
correct weight for your specic model.
- English