06 Preset Single Register
Presets a value into a single register of the Model 2000 analyzer.
Query
The query message specifies the register to be preset. The demand clean cycle register is the only
register in the Model 2000 which can be written to. When any value is written to this register, a clean
cycle is initiated. The address of the demand clean cycle register is 238C (hex).
Below is an example of a request for a demand clean cycle on an analyzer with the slave address of
1.
Field Name Example
Slave Address 01
Function 06
Reg. Address Hi 23
Reg. Address Lo 8C
Data Hi 00
Data Lo 00
CRC --
The normal response is an echo of the query.