Loading...
3
Contents
Perfusor fm
®
Overview page 4
Operation page 6
Special Functions page 8
Reasons for Alarm page 12
Cleaning / Battery page 13
Start-up graphs and trumpet curves page 14
Technical data page 16
Warranty page 18
Order Data page 19
The Perfusor fm is a transportable infusion
syringe pump in accordance with IEC/EN
60601-2-24 that is suitable for dispensing
liquids in nutritional and infusion therapy. The
medical specialist must decide on suitability
for application on the basis of the warranted
properties and the technical data.
For further details please refer to these
Instructions for Use.
The Perfusor fm is used primarily for the highly
precise, reproducible infusion of medication in
large and small quantities. The device permits
extremely exacting delivery rate settings, and
thus optimal adaptation to any given
requirements.
The drive system ensures reliable gripping of
the syringe on pressing the appropriate key,
which in turn makes for optimal start-up and
considerably facilitates handling when
inserting the syringe.
9-stage pressure limitation provides extra
safety. Pressure values and trends are
displayed during infusion. In the event of a
pressure alarm, the bolus volume built up by
the Perfusor fm is automatically reduced. In
addition, operation of the unit can be made
even more reliable if required with double
entry of the rate.
Special functions broaden the scope of
applications. Additional memory cards can
provide yet more functions, for example
conversion to patient-controlled analgesia
(PCA).