Additional settings
To Change the Delivery Rate
Press STOP.
Press C. Display is reset to 000.0.
Enter new delivery rate.
Press START. Infusion is implemented.
(Alarm is given if there is no restart
within 2 minutes)
To Change the Rate Without Interrupting the
infusion (Function can be deactivated by our
Service staff)
Enter new rate.
Press key beneath RATE. Rate is transferred
to red display. Unit operates at new rate. (If
new rate is not confirmed within 10 s, infusion
continues at old rate)
Target Volume (Volume Pre-selection)
The target volume will be administered inde-
pendent on the infused volume.
Press key below VOL.
Enter target volume via keypad and confirm
with VOL (values between 0.1 and 999.9 ml).
After confirming the display shows the residual
volume instead of the target volume.
Note: When target-/residual volume has been
administered, the pump stops and goes into
alarm. Acknowledge alarm and press VOL. Clear
displayed residual volume with C (in display
appears ----,- ml). Confirm this with VOL or
enter new target volume. To continue infusion,
the target volume has to be displayed as ---,-
ml or a new target volume has to be set.
Target Time (Time Pre-selection)
Press key beneath TIME.
Enter target time via keypad and confirm with
TIME (50 min. = 5 0; 2 h/30 min. = 2 3 0). After
confirming instead of the target time the resi-
dual time is displayed.
Correction: Press C. Display --h --m.
New entry.
Note: When time has been expired, the pump
stops and goes into alarm. Acknowledge alarm
and press TIME. Clear displayed residual time
with C (in display appears --h -- m). Confirm
this with TIME or enter new target time.
Rate calculation
Displayed delivery rate must be 000.
0.
Enter volume and time. The delivery rate
is calculated automatically and displayed
(rounded to one decimal place).
Confirm RATE.
Press START. Infusion is implemented.
Clear Time/Volume
Press STOP, then key beneath VOL or TIME.
Press C to clear.
1x: Target Vol/Time = ----.-ml / --h--m
2x: Infused Volume/Real Time = 0.0ml /00h00m
Status Request
Only when infusion is running.
Press 1 x resp. 2 x key below INFO for
actual values. The display disappears after 10
sec or after all information has been requested.
Press x times key below INFO for displaying
syringe type, battery capacity in h and min as
well as hours of operation.
Set rate is automatically reduced if required
for type of syringe being used.Make sure syringe is properly inserted