Configure jumpers P2-P8 for desired output and connect terminals on TB5 for voltage or current.
Ensure all S1 switches are in the OFF position for proper operation of the analog output.
Navigate through the indicator's menu using specific key presses to access setup options.
Enable and configure the analog output (AN.OUT) by setting LO and HI values for the output range.
Perform ZERO calibration before SPAN, adjusting output to be slightly above 4.00 mA or 0V.
Adjust SPAN calibration to be slightly below 20.00 mA or 10V, then re-check ZERO if needed.
Set up a protocol (PROT 1-5) with TYPE=B-CAST, BIND=OPTION, ATTR settings for continuous output.
Save the configured protocol settings and initiate a reboot to apply changes.