Instruction for use
26
En
PRO-33 / PRO-35
The PRO-33 / PRO-35 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the PRO-33 / PRO-35 can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the PRO-33 / PRO-35 as recommended below, ac-
cording to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter,
m
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
800 MHz to 2.5 GHz
d = 2.3 P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and re ection from
structures, objects and people.
14. LATEST REVISION
Latest revision: 2016-08-10
Exclusive distributor in Pakistan:
IHI - INTEGRATED HUMAN INITIATIVES
PO Box # 12301, Post Code 75500
DHA, Karachi, Pakistan.
Mobile: 0300 927 4 927
Telephone: 021 3585 7973
Email: HYPERLINK “mailto:info@ihumani.com” info@ihumani.com
Website: HYPERLINK “http://www.ihumani.com” www.ihumani.com

Related product manuals