Instruction 0011-9026 Page 9
Figure 12. When plunger
valve is depressed, a gas
sample is pumped through
top reservoir with center
bore is sealed off.
Figure 13. When plunger
valve is released, the gas
sample is locked into FY-
RITE and the top reservoir
is opened to center bore so
that gas sample can pass
through absorbing uid.
Figure 14. Absorption of
gas sample by uid cre-
ates suction, which causes
diaphragm to ex up and
uid to rise in center bore
to replace gas absorbed.
4.0 PRINCIPLE OF OPERATION
(Refer to Figures 12, 13, and 14)