B&C Electronics PH 7685/E – Rev. B
- 28 -
Set-point adjustment
SA: Set-point A adjustment
14.00pH (600mV): actual Set-point value
ù to stop the procedure and to go back to ·D3· (·D4·)
3.ò ç to insert the Set-point value
4.£ to confirm and to go to the next step
5.3.5 On/Off function
0.15pH (15mV): actual hysteresis value
ù to stop the procedure and to go back to ·D3· (·D4·)
1.ò ç to insert the hysteresis value
2.£ to confirm and to go to the delay time selection
5.0 s: actual delay time value
ù to stop the procedure and to go back to ·D3· (·D4·)
CAL SA S:14.00pH
CAL SA S: 600mV
CAL SA I: 0.15pH
CAL SA I: 15mV
CAL SA D: 5.0s