EasyManua.ls Logo

Bard PowerPICC Solo2 - Placement Precautions and Complications

Default Icon
10 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
Some patients may be hypersensitive to heparin or suer from heparin induced thrombocytopenia (HIT) and these patients
must not have their catheter locked with heparin ush solution.
As reported in literature, anaphylactic or anaphylactic-like reactions occur in a small percentage of the population during
placement
1
, positioning
1
, ushing
2
of central venous catheters or cleaning of catheter exit site
3
. These reactions are reported
in association with insertion, rapid ushing, or manipulation of the catheter and/or use of chlorhexidine gluconate (CHG) in
some patients. Be aware of the potential symptoms or signs of these reactions and take precautionary steps as dictated by
institution protocol for their prevention or treatment.
If CHG allergy is suspected, conrmatory testing is recommended
4,5
.
Precautions Related to Device Placement Procedure
The PowerPICCSOLO*
2
catheter features a reverse-taper catheter design. Placement of larger catheters at or below antecubital
fossa may result in an increased incidence of phlebitis. Placement of the PowerPICC SOLO
2
® catheter above antecubital fossa
is recommended.
Avoid placement or securement of the catheter where kinking may occur, to minimize stress on the catheter, patency
problems or patient discomfort.
Flush the catheter with sterile normal saline prior to use. Catheter stylet must be wetted prior to stylet repositioning or
withdrawal.
Do not advance the guidewire past the axilla without uoroscopic guidance or other tip locating methods.
If the guidewire must be withdrawn while the needle is inserted, remove both the needle and wire as a unit to prevent the
needle from damaging or shearing the guidewire.
Never use force to remove the stylet. Resistance can damage the catheter. If resistance or bunching of the catheter is
observed, stop stylet withdrawal and allow the catheter to return to normal shape. Withdraw both the catheter and stylet
together approximately 2 cm and reattempt stylet removal. Repeat this procedure until the stylet is easily removed. Once the
stylet is out, advance the catheter into the desired position.
Avoid accidental device contact with sharp instruments and mechanical damage to the catheter material. Use only smooth-
edged atraumatic clamps or forceps.
Avoid perforating, tearing or fracturing the catheter when using a guidewire.
Avoid sharp or acute angles during implantation which could compromise the patency of the catheter lumen.
The PowerPICC SOLO
2
® catheter is designed for use with needleless injection caps or direct-to-hub” connection technique.
Apply a sterile end cap on the catheter hub to prevent contamination when not in use. Use of a needle longer than 1.6 cm
may cause damage to the valve.
Do not use scissors to remove dressing to minimize the risk of cutting catheter.
Do not suture through or around any part of the catheter’s tubing (shaft or extension legs). If using sutures to secure catheter
USE THE SUTURE WINGS and make sure they do not occlude, puncture, or cut the catheter.
The catheter must be secured in place to minimize risk of catheter breakage and embolization.
To reduce potential for blood backow into the catheter tip, always remove needles or syringes slowly while injecting the last
0.5 mL of saline.
Do not withdraw dilator from microintroducer sheath until sheath is within vessel to minimize the risk of damage to sheath
tip.
Do not pull apart the portion of the sheath that remains in the vessel. To avoid vessel damage, pull back the sheath as far as
possible and tear the sheath only a few centimeters at a time.
Do not cut guidewire to alter length.
Do not insert sti end of guidewire into vessel as this may result in vessel damage.
Keep sucient guidewire length exposed at hub to allow for proper handling. A non-controlled guidewire can lead to wire
embolism.
Do not use excessive force when introducing guidewire or microintroducer as this can lead to vessel perforation and bleeding.
Never leave stylet or stiening wire in place after catheter insertion; injury may occur. Remove stylet or stiening wire and
T-lock (as applicable) after insertion.
The stylet or stiening wire needs to be well behind the point the catheter is to be cut. NEVER cut the stylet or stiening wire.
Possible Complications
The potential exists for serious complications including the following:
Air Embolism
• Bleeding
Brachial Plexus Injury
Cardiac Arrhythmia
Cardiac Tamponade
Catheter Erosion Through the Skin
Catheter Embolism
Catheter Occlusion
Catheter Related Sepsis
• Endocarditis
Exit Site Infection
Exit Site Necrosis
• Extravasation
Fibrin Sheath Formation
• Hematoma
Heparin Induced
Thrombocytopenia
Hypersensitivity, anaphylactic or
anaphylactic-like reaction during
placement
1
, positioning
1
, ushing
2
of catheter or cleaning of catheter
exit site
3
Intolerance Reaction to Implanted
Device
Laceration of Vessels or Viscus
Myocardial Erosion
Perforation of Vessels or Viscus
• Phlebitis
Spontaneous Catheter Tip
Malposition or Retraction
• Thromboembolism
Venous Thrombosis
Vessel Erosion
Risks Normally Associated with
Local or General Anesthesia,
Surgery, and Post-Operative
Recovery

Related product manuals