Section 3. Installation
Inspection of Shipment After carefully unpacking the instrument, check the contents of the packages and inspect for
damage. Parts of the C3 Cell Stand are listed below. (This list is subject to change.) Please
refer to the packing slip with your instrument to verify the parts. Retain the shipping box and
packing material until you have fully tested the unit to be certain that no damage was
incurred during shipment.
If you discover a shortage in parts, call BAS Customer Service and describe the shortage.
Replacement part(s) will be sent immediately, subject to stock availability.
Parts of C3 Cell Stand C3 Cell Stand with Faraday Cage
Stir bar
Glassy carbon voltammetry electrode
Platinum voltammetry electrode
Voltammetry cell vials
Polishing kit
Ag/AgCl reference electrodes (3)
Platinum wire auxiliary electrode
Port plug
Gas connection tubing
Cables
Instruction manual
Location of C3 Cell Stand 1. Provide a surge-free power source. Other laboratory equipment such as ovens, vortex mix-
ers, centrifuges, and large motors may cause spikes in the power supply.
2. Ensure that all components of the system share the same ground circuit. This can best be
accomplished by plugging all components into a multi-outlet power strip. Plugging the
components into independent outlets can produce ground loops (current that flows be-
tween ground circuits at slightly different potentials), which can produce baseline noise.