Quick-Reference Guide for using the ePG together with other Beamex products - 1
Quick-Reference Guide for using
the ePG together with other
Beamex products
Introduction
This document provides an overview on how to use your MC6 family calibrator
to control the ePG Electric Pressure Pump and Controller in your daily work.
It also describes what to consider when working with Beamex Calibration
Management Software (CMX or LOGiCAL).
Requirements
• MC6 family calibrator (MC6, MC6-WS, MC6-Ex or MC6-T) with firmware
version 4.30 or later and “Pressure controller communication, ePG” option
installed on the calibrator.
Note: Both can be checked in the calibrator’s Settings > About
window.
• ePG firmware version 2.00 or later.
Enable communication between an MC6
family calibrator and the ePG
MC6 family calibrator does not automatically recognize your ePG. You need to
configure the controller communication to pair the devices.
Note: If you have more than one ePG, you can assign each ePG on
a separate Controller Presets channel (maximum 4).