7. BERCHTOLD has tested quaternary disinfectants for compatibility with
the hand control materials, and approves their use.
8. BERCHTOLD contraindicates the use of disinfectants that contain so-
dium hypochlorite on hand controls. E.g., Clorox Health-care Dis-
patch.
9. BERCHTOLD cannot recommend the use of other types of disinfec-
tants on the hand controls.
10. Allow hand controls to fully dry after disinfectant manufacturer’s rec-
ommended “set time” before re-cleaning.
11. Do not submerge the hand control in any fluids.
12. If the hand control is submerged, remove the hand control from serv-
ice immediately. Contact BERCHTOLD technical service for evalua-
tion assistance
13. Do not use abrasive cleaners (solutions with pumice, scotch bright
pads, etc…) when cleaning and disinfecting hand controls.
14. Before each use, users should inspect the hand control for signs of
damage or abuse (e.g. frayed cable, cracked housing, missing fas-
teners, lack of glue beak on face. etc…)
15. If damage is suspected, remove from service immediately and con-
tact BERCHTOLD technical service.
16. The maintenance activity of the facility should put hand controls on a
monthly PM schedule for a thorough inspection.
17. The following provide areas for inspection and key points for inspec-
tion and cleaning of hand controls
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Operation
59