55
If the volume control is activated
A warning may be displayed if you activate the
tinnitus volume control in the “Local Controls
screen. This occurs if the relief sound can be listened
to at levels that may cause hearing damage. The
Limitation on Use” table in the tting software
displays the number of hours the patient can safely
use Tinnitus SoundSupport.
· Note the max wearing time for each program for
which Tinnitus SoundSupport is activated.
· Write those values in the table: “Tinnitus
SoundSupport: Limitation on use, in the back of
this booklet.
· Instruct your patient accordingly.