6
In the switch on state, press "MEM" button and hold for 5 seconds, and all data in memory
will be deleted. Then press "MEM” button again, the display will show "CLr".
5. High-temperature prompt function
Under the body temperature mode, if the measured body temperature is higher than or equal
to 38
, the product will beep continuously.
6. Alarm setup
When the device is on, Press “MODE” button for 2 seconds, the screen will display F1”, then
press “MEM” button once, the screen will display “F2”, press “MODE” button choose the
alarming temperature from 37.3°C to 39.1°C(99.1°F to 102.4°F).Confirm by pressing MEM"
button.
Note: The alarm threshold default value is 38°C (100.4°F)
7. When the device is on, press to turn the buzzer on or off.
When the screen shows “ON”, the buzzer opened.
When the screen shows “OFF”, the buzzer closed.
VII. TECHNICAL SPECIFICATIONS
Items
Specifications
Battery
DC 3V 2 pcs AA dry batteries
Temperature sensor
Infrared Sensor
Measuring distance
3cm - 5cm(1.2in - 2in)
Measurement range
Body temperature mode:
32.0°C ~ 43.0°C (89.6°F ~ 109.4°F)
Under body mode, there is three color backlits:
Green color backlit: less than 37.3°C (99.1°F), means
normal temperature.
Orange color backlit: 37.4°C - 37.9°C (99.3°F - 100.2°F),
means may be low fever.
Red color backlit: more than 38°C (100.4°F), means
may be high fever.
Surface temperature mode:
0°C ~ 60°C (32°F ~ 140°F)
Room temperature mode:
0°C ~ 40°C (32°F ~ 104°F)
Measuring accuracy
Body temperature mode:
32.0°C ~ 34.9°C (89.6°F ~ 94.8°F) : ±0.3°C (±0.6°F)
35.0°C ~ 42.0°C (95°F ~ 107.6°F) : ±0.2°C (±0.4°F)
42.1°C ~ 43.0°C (107.8°F ~ 109.4°F) : ±0.3°C (±0.6°F)
Clinical repeatability
≤0.2°C (0.4°F)
Display resolution
0.1°C (0.1°F)
Measurement Time
1 second
Memory
32 sets
Operation Environment
Ambient temperature: 10°C ~ 40°C (50°F ~ 104°F)
Relative humidity: ≤85%RH
Atmospheric Pressure: 700 hPa to 1060 hPa
Storage and transportation condition
Ambient temperature: -20 ~ 55 (-4 ~ 131)
Relative humidity: 95%RH
Atmospheric Pressure: 700 hPa to 1060 hPa
Weight
Approx.80g (Battery excluded)