In the switch on state, press "MEM" button and hold for 5 seconds, and all data in memory
will be deleted. Then press "MEM” button again, the display will show "CLr".
5. High-temperature prompt function
Under the body temperature mode, if the measured body temperature is higher than or equal
to 38
℃
, the product will beep continuously.
6. Alarm setup
When the device is on, Press “MODE” button for 2 seconds, the screen will display “F1”, then
press “MEM” button once, the screen will display “F2”, press “MODE” button choose the
alarming temperature from 37.3°C to 39.1°C(99.1°F to 102.4°F).Confirm by pressing “MEM"
button.
Note: The alarm threshold default value is 38°C (100.4°F)
7. When the device is on, press “ ” to turn the buzzer on or off.
When the screen shows “ON”, the buzzer opened.
When the screen shows “OFF”, the buzzer closed.
VII. TECHNICAL SPECIFICATIONS
DC 3V 2 pcs AA dry batteries
Body temperature mode:
32.0°C ~ 43.0°C (89.6°F ~ 109.4°F)
Under body mode, there is three color backlits:
Green color backlit: less than 37.3°C (99.1°F), means
normal temperature.
Orange color backlit: 37.4°C - 37.9°C (99.3°F - 100.2°F),
means may be low fever.
Red color backlit: more than 38°C (100.4°F), means
may be high fever.
Surface temperature mode:
0°C ~ 60°C (32°F ~ 140°F)
Room temperature mode:
0°C ~ 40°C (32°F ~ 104°F)
Body temperature mode:
32.0°C ~ 34.9°C (89.6°F ~ 94.8°F) : ±0.3°C (±0.6°F)
35.0°C ~ 42.0°C (95°F ~ 107.6°F) : ±0.2°C (±0.4°F)
42.1°C ~ 43.0°C (107.8°F ~ 109.4°F) : ±0.3°C (±0.6°F)
Ambient temperature: 10°C ~ 40°C (50°F ~ 104°F)
Relative humidity: ≤85%RH
Atmospheric Pressure: 700 hPa to 1060 hPa
Storage and transportation condition
Ambient temperature: -20℃ ~ 55℃ (-4℉ ~ 131℉)
Relative humidity: ≤ 95%RH
Atmospheric Pressure: 700 hPa to 1060 hPa
Approx.80g (Battery excluded)