Do not press the buttons with excessive force or with
pointed objects.
The monitor can maintain the safety and performance
characteristics for a minimum of 10.000 measurements
or 3 years
NOTE:
Further documents which include information for the Cali-
bration of the device can be asked at the service address.
The device can be tested in case of supposed inaccur-
acy every two years. Please contact the nearest service
address.
To keep your digital blood pressure monitor in the best con-
dition and protect the unit from damage follow the direc-
tions listed below:
Clean your device and cuff carefully only with a slightly
moistened soft cloth and dry it immediately with a soft
dry cloth. Do not press.
When storing the device, make sure that no heavy objects
are placed on top of it.
Do not fold the cuff and tubing tightly. The cuff tube
should not have any sharp kinks and keep it away from
sharp edges.
If the unit is stored near freezing, allow it to acclimate to
room temperature before use.
Always store the unit in its storage bag after use.
This unit may not meet its performance specifications if
stored or used outside of the temperature and humidity
ranges (see section “10. Technical Specifications”).
Use the unit consistent with the instruction provided in
this manual.
9. Error messages / troubleshooting
In the event of errors, the error message _ appears on
the display.
Error messages may appear if:
systolic or diastolic pressure could not be measured (
or appears on the display),
systolic or diastolic pressure was outside the measure-
ment range ( or Lo appears on the display),
the cuff is fastened too tightly or loosely ( or ap-
pears on the display),
the pump pressure is higher than 300 mmHg ( ap-
pears on the display),
pumping up takes longer than 160 seconds ( appears
on the display),
there is a system or device error ( , , or
appears on the display),
the batteries are almost empty .
In such cases, repeat the measurement. Ensure that you
do not move or speak.
If necessary, reinsert or replace the batteries.