32 Beurer GL40 mmol/L
8 Troubleshooting
Messages on the display for batteries and blood sugar measurement
No. Cause Action
LP Flat batteries Replace all batteries.
Ht
Temperature of the measure-
ment room, measuring device
or test strip above the allow-
able limit.
Repeat test with new test strip as soon as the
measurement room, measuring device and test
strip have reached room temperature (+20°C to
+25°C).
Lt
Temperature of the measure-
ment room, measuring device
or test strip below the allow-
able limit.
Repeat test with new test strip as soon as the
measurement room, measuring device and test
strip have reached room temperature (+20°C to
+25°C).
Err Used or contaminated
test strip was inserted.
Insert test strip which has not been used or
spoiled.
Repeat blood sugar measurement.
001 System error
Remove the batteries and reinsert them.
Contact our customer service department if
the problem persists.
005 Memory error
Remove the batteries and reinsert them.
Contact our customer service department if
the problem persists.