Beurer GL50 mmol/L 37
Usage limits for specialist personnel from the healthcare sector:
1
. If the patient exhibits the following symptoms, it may be the case that no correct values can be ob-
tained:
•Acutedehydration
•Acutehypotension(lowbloodpressure)
•Shock
•Hyperosmolarhyperglycaemiccondition(withorwithoutketosis)
2
. Lipaemic samples: cholesterol levels up to 13 mmol/L and triglycerid levels up to 11.4 mmol/L do
not influence the results. Severely lipaemic blood samples were not tested with the Beurer GL50
monitor; therefore, using the device with these samples is not recommended.
3
. Blood glucose monitors designed for home use should not be used for severely ill patients.
4
. The eect of interfering substances on the measurements depends on the concentration in the blood.
The maximum concentrations of certain substances listed below do not significantly influence the
measurements.
Blood glucose value
4.4 mmol/L 6.7 mmol/L 27.8 mmol/L
Concentration of
tested substances
Ascorbic acid
0.23 mmol/L
10.9% -1.8% 4.6%
Ibuprofen
2.42 mmol/L
3.1% 2.9% 4.6%
L-dopa
0.09 mmol/L
10.6% 7.9% 4.9%
Sodium salicylate
3.12 mmol/L
-2.6% 9.4% -0.8%
Tetracyclin
0.03 mmol/L
-5.3% 3.8% 3.2%
Tolbutamide
3.70 mmol/L
-2.6% 12.3% 0.9%
Unconjugated bilirubin
0.04 mmol/L
-2.5% 4.1% -0.2%
Uric acid
0.48 mmol/L
2.7% 9.6% -1.8%
Xylose
0.27 mmol/L
7.6% 0.8% 7.5%
Influence