AutoTune
1:)Plug on
220Volts 50/60Hz Open
the power switch.
2:)Press "set" then increase or decrease to target temp (eg.
Sp=33).
then press "set" (St=0) and then press "set"
6 seconds
save and exit.
It is easily to operate in default leaving factory for
purging drying process.
If client are not
satisfied with target temp (always find over heat problem)during heat process.
How to make it normal working? Easily Follow below steps:
1:)Power Off Vac Oven and Open Door Cool down inside oven temp below 33deg C----(Cool Down 30-50minutes).
2:)Power On.Then Press "set" 6" it will showing Lc" increase 3 then entry into "Menu I" Press "Set" again to find out "P"
increase "35" press "set" again I" increase "200" press "set" then D" increase "200" then press "set" 6" save and exit.
3:)Press "AT" 6" it will entry into Autotune Process for (eg
33deg C ) The "AT" light flashing and working to find out the best
PID data and save and exit
(once it finished the light always light on).Then Power Off. Wait 30 minutes.
Autotune Process
function:in order to avoid Over Heating Temp(eg. You want 48deg C if AT,it will not over heat higher than
<50deg C) to protect material.
Caution:Autotune Process 50deg C ( 30-50deg C ) the uniform temp working well and without over heat temp.