Key guidelines for installing the ExDetector device in a safe environment.
Critical safety advice for making electrical connections to the device.
Details the specific wiring connections for the detector terminals.
Illustrates how to connect the ExDetector to a GMC8022.
Necessity of connecting housing to equipotential bonding in hazardous zones.
How to connect cable screen to housing for interference reduction.
Procedures for calibrating detector settings using test gases and voltmeter.
Steps to adjust bridge voltage to 0 mV using synthetic air.
Instructions to set the base current to 4 mA using the N potentiometer.
How to adjust the V potentiometer for the 4-20 mA output signal.
Procedure to measure and record bridge voltage for sensitivity checks.
Formula and criteria for calculating and evaluating sensor sensitivity.
Step-by-step guide to safely replace the sensor unit of the ExDetector.
| Operating Temperature | -20°C to +50°C |
|---|---|
| Operating Humidity | 15% to 90% RH (non-condensing) |
| Power Supply | 3.6 V Lithium battery |
| Output Signal | 4-20 mA |
| Ingress Protection | IP65 |
| Certifications | ATEX, IECEx |
| Protection Class | IP65 |