Dental Handpiece Maintenance Device
Distributed by: Bien Air Dental SA
Manufactured by: J.MORITA MFG. CORP.
2021-05-21 Pub. No.: K345-91001-503 (EN) Printed in Japan
Lubricare 2
INSTRUCTIONS FOR USE

Questions and Answers

  • T
    Timothy CoxAug 8, 2025
    What to do if my Bien Air Dental equipment process stops when pressing the Start/Stop switch?
    • D
      Donna GrantAug 8, 2025
      If the process stops when you press the Start/Stop switch, all Line Selection Lamps blink, and a triple beep sounds, it could be due to several reasons: * The front door might be open. Ensure that the front door is properly closed and press the Start/Stop Switch again. * The front door might be dislocated, or the hinge might not be inserted properly. Remove the front door and reinstall it, ensuring the hinge is inserted straight until it snaps on. * The door seal might be partially removed. Reinstall the door seal properly.
  • A
    Ashley Irwin MDAug 12, 2025
    Why does my Bien Air Lubricare 2 HIM-2 release oil mist and make installation surfaces oily?
    • J
      Judy HansenAug 12, 2025
      Oil mist release and oily installation surfaces can occur due to several factors: * The handpiece shape may cause more oil scattering during maintenance. Reduce the air pressure to 0.3 MPa. * The handpiece might be attached in an improper direction. Ensure the handpiece direction is correctly set. * The main unit’s air pressure setting may be too high. Adjust the regulator and set the pressure within the specified range. * If using a straight attachment, a bur might not be inserted. Refer to the user’s manual that comes with the handpiece, as some handpieces may release oil mist if a bur is not inserted.
  • C
    Curtis KerrAug 16, 2025
    What to do if my Bien Air Dental equipment process stops before lubrication?
    • B
      Brent MasonAug 16, 2025
      If the process stops before lubrication, the selected Line Selection Lamp and corresponding Empty Spray Can Lamp blink, and a short series of beeps sounds, it could be due to these reasons: * The spray can might be empty. Replace the spray can with a new one. * The spray can might not be properly installed. Fix the spray can and spray stand installation. * Compressed air might not be supplied. Connect the air tube correctly. * The air regulator setting might not be within the correct range. Adjust the regulator and set the pressure within the specified range. * The air tube might be bent. Fix the air tube.
  • C
    Carlos AcostaAug 20, 2025
    Why does my Bien Air Lubricare 2 HIM-2 Dental equipment have an odd odor and get unusually dirty?
    • E
      Eduardo AdamsAug 20, 2025
      An odd odor and unusual dirtiness in the device can be caused by: * A handpiece was not connected to the lubrication line. Ensure only lines with handpieces connected are selected. * The oil absorbent pad is saturated. Remove the old pad, clean the device, and set a new oil absorbent pad. * The oil absorbent sheet is dirty. Remove the old sheet, clean the device, and set a new oil absorbent sheet. * The spray can is not properly installed. Fix the spray can and spray stand installation. * Air pressure is set out of the specified range. Adjust the regulator and set the pressure within the specified range.
  • M
    Michael FlowersAug 25, 2025
    How to fix excessive oil seepage from my Bien Air Lubricare 2 HIM-2 Dental equipment?
    • R
      rebecca27Aug 25, 2025
      Excessive oil seepage from the connection parts and the handpiece head during normal use can be due to: * Lubrication was not performed in the recommended mode, or chuck lubrication was performed on a handpiece that does not require it. Refer to the user manual that comes with the handpiece for the recommended mode. * The main unit’s air pressure setting is too low. Adjust the regulator and set the pressure within the specified range. * The handpiece was stored with its head down. Store the handpiece in an upright position with its head at the top. * The handpiece was used without wiping excess oil after maintenance. Wipe excess oil from the handpiece with a soft cloth or gauze after maintenance, being careful not to damage the handpiece surface.
  • M
    Monica SmallAug 29, 2025
    What to do if my Bien Air Lubricare 2 HIM-2 Dental equipment has no power?
    • M
      Matthew JacobsonAug 29, 2025
      If there is no power when the Power Switch is turned on, it might be because the power cord is not properly plugged into the power outlet. Connect the power cord correctly.
  • K
    Kathy SimpsonSep 2, 2025
    Why lubrication does not start when pressing the Start/Stop Switch on my Bien Air Lubricare 2 HIM-2?
    • A
      Anne JohnsonSep 2, 2025
      If lubrication does not start when the Start/Stop Switch is pressed, it could be because no line is selected. Select at least one line before pressing the Start/ Stop Switch.