Garment Cleaning and Disinfecng—connued from previous page
4. Garment should be soaked for 30 minutes with mild agitaon every 5 to 10 minutes while
keeping it below water surface.
5. Thoroughly rinse garment with warm tap water and allow to air dry.
WARNING! Never allow the Latch Connectors to be submerged into the water. If water
enters the inside of the garment, damage may occur to the device.
6. Harder to remove soil on surface of garment may require addional washing by hand with a
clean towel while submerged. Avoid using any abrasive materials such as scrubbing pads or
chemicals that could cause damage to the exterior surface of garment.
7. Re-Submerge garment for 30 minutes (with excepon of tubing connectors) in soluon con-
sisng of 1 cup of bleach per 1 gallon of warm tap water, again agitang garment every 5 to
10 minutes while keeping garment below water surface. Rinse garment thoroughly with
warm tap water and allow to air dry. This completes the disinfecng step.
WARNING! DO NOT place garment in washing machine.
WARNING! DO NOT use the tubing or valves as “handles” for carrying, handing or
storing garment.
TROUBLESHOOTING
If the correcve acon does not resolve the problem, please call Bio Compression Systems, Inc., at
1-800-888-0908. Serial number is required when calling for service.
Symptom Possible Cause Correcve Acon
The device is not working.
No electricity Check the electrical wall outlet to be sure that
the pump is plugged into the outlet correctly.
Check the circuit breaker to be sure there is
power to the outlet.
Power cord
Unplug the power cord and look for any
damage or defects.
One garment inates but the
second one does not.
The second garment is not
receiving air.
Check the garment hoses for adequate con-
necon to the device, kinks, punctures, twists
and /or folds.
The device is making strange
and/or loud noises.
Device is on a uneven or unsta-
ble surface
Move to a more stable surface.
An internal problem Contact Bio Compression Systems, Inc., for
repair.
Regardless of the pressure
seng the garments are
applying a very low pressure.
Defecve Garment Check the garment for adequate connecon
to the device, leaks, kinks, punctures, twists
and /or folds.
An internal problem Contact Bio Compression Systems, Inc., for
repair.