CAUTION: In case of power failure, the Sequenal Circulator shall automacally power
o. The pump does not have a baery backup nor mer memory. Once power is re-
stored, the Sequenal Circulator will need to be powered on in order to resume treat-
ment. User must keep apprised of alloed treatment me.
GENERAL GUIDELINES FOR TREATMENT
Physician must prescribe sengs for use with this device. The general guidelines below
are for informaonal purposes only and should not replace sengs provided by
physician. 60mmHg is an eecve seng for many paents. Physician must provide
paent’s sengs.
The presence of broc ssue may require a pressure of 80mmHg in order to soen the
broc ssue and achieve reducon. Once ssue soens, compression may be reduced
to 60mmHg, if determined by physician
Paents with a history of Congesve Heart Failure (CHF), which is controlled by
medicaon, should never be in a at posion while using the Sequenal Circulator. These
paents should be in a reclined posion with their legs elevated during treatment. It is
suggested that treatment regimens be divided into two daily 30-minute treatments.
Paents with a history of Deep Vein Thrombosis (DVT), with or without a lter, may
require less compression pressure, 40mmHG is generally tolerable but must be
prescribed by a physician. Paents with a lter may need to divide treatment into two
daily 30-minute sessions. It is suggested that the provider obtain a Negave Doppler
Study from the physician for their records.
MAINTENANCE AND STORAGE
Exterior Pump Case Cleaning instrucons:
1. Clean the exterior case and tubing with a damp (not wet) cloth using mild soap and
water soluon once per month
WARNING!
Only an authorized technician may open the pump
Before cleaning, unplug power cord from electrical outlet
GARMENT CLEANING/DISINFECTING INSTRUCTIONS:
Disconnect garment from device.
2. Open garment to expose all sides either by separang Velcro type hook and loop or
by unzipping (depending on type of garment).
WARNING!
Do not allow liquids to enter pump, as this can present an electrical hazard
Always allow the pump to dry before using
Do not use bleach on the pump
3. Cleaning soluon should consist of 1/3 cup of laundry detergent per 1 gallon of
warm tap water. Use either a large sink or plasc tub able to hold enough soluon
(depending on size and quanty of garments) to completely submerge the garment
leaving the latch connector bars out of the water.