EasyManua.ls Logo

Bio Compression Systems SC-2004-OC - Operating Instructions; Device Preparation Steps; Connecting Garment Tubing

Bio Compression Systems SC-2004-OC
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Powering On the Sequenal Circulator
11. Paent shall be seated in a reclined posion, legs elevated within easy reach of the
Sequenal Circulator.
12. Press the “Power On/O” buon to turn the pump on.
13. Within several seconds, the pump shall begin producing air pressure. The air pressure
may be set in the same posion as the previous treatment. Factory default pressure is
50 mmHg in the chamber #1, decreasing incrementally as it moves distally to proximally
between zones.
Seng the Air Pressure on the Sequenal Circulator
14. Turn Pressure Knob to the prescribed seng.
15. The Pressure Knob determines the pressure in the distal chamber. Once set, the
Sequenal Circulator shall automacally set the subsequent pressure zones within the
garment, by incrementally decreasing each zone as the air pressure moves distally to
proximally.
16. Treatment me shall be determined by physician. Adhere to physician’s instrucon on
me alloed for each treatment.
Powering O the Sequenal Circulator
17. Upon compleon of treatment, press the “Power On/O” buon to turn o the pump.
The pump shall stop running at this me. Please note you must wait unl the light goes
out before removing your garment.
18. Prior to removing the garment, squeeze the sides of the Latch Connector Bars and
detach from pump by pulling outward away from pump. This will fully detach the
garment from the pump.
19. Depress the garment to help evacuate any remaining air. Once the garment becomes
loose enough to remove, unzip the garment and remove.
20. Repeat removal procedure for any addional garments being used.