Questions and Answers

  • K
    kelsey93Aug 13, 2025
    How to fix BIO RAD PowerPac 200 E 1 No Load Detected error?
    • A
      Ashley ParksAug 13, 2025
      If your BIO RAD Power Supply shows error code E 1, which means no load is detected, verify all electrical connections to the electrophoresis cell. Also, ensure the cell contains the correct buffer volume. Once checked, press the key to resume the run, or press any other key to clear the error code.
  • M
    Miguel LynchAug 17, 2025
    What to do if BIO RAD PowerPac 200 shows E 2 Short Circuit error?
    • D
      Daniel GrahamAug 17, 2025
      If the BIO RAD Power Supply displays error E 2, indicating a short circuit, it means the current load has exceeded safe limits (2.5 A for PowerPac 200 or 400 mA for PowerPac 300). You should inspect the setup for any short circuits or excessive loads. If the high load is due to a high buffer concentration, remake the buffer. After correcting the issue, press the key to resume the run or any other key to clear the error code.
  • R
    regina70Aug 20, 2025
    What to do when BIO RAD PowerPac 200 displays E 10 Unacceptable Value(s) Entered?
    • D
      Derrick GonzalezAug 21, 2025
      If the BIO RAD Power Supply shows error E 10, indicating unacceptable values were entered, clear the error by pressing any key (except the key). Make sure to enter acceptable values for voltage and current. Note that for PowerPac 200 the voltage must be at least 5 V and the current at least 0.01 A, while for PowerPac 300 the voltage must be at least 10 V and the current at least 4 mA. After entering correct values, press the key to resume the run.
  • K
    kristenfrazierAug 25, 2025
    What does E 3 Power Supply Regulation Error mean on BIO RAD?
    • P
      patkinsonAug 26, 2025
      If the BIO RAD Power Supply displays error E 3, indicating a power supply regulation error, check for issues such as dirty contacts, frayed wires, or excessive buffer concentration and correct them. Then, press the key to resume the run or press any other key to clear the error code. If the error persists, note the error code number and contact your Bio-Rad Representative.
  • B
    Brittany ReidAug 28, 2025
    How to resolve E 11 Ground Leak Detected error on BIO RAD PowerPac 200?
    • B
      Barbara GreenAug 28, 2025
      If the BIO RAD Power Supply (PowerPac 200 only) displays error E 11, which means a ground leak was detected, inspect the cell for an improper connection to earth ground and correct it. Then, restart the power supply by turning it off and then on again.
  • C
    carolyn09Sep 1, 2025
    What does E 5 Power Failure During a Timed Run mean on BIO RAD PowerPac 200?
    • R
      Renee HernandezSep 1, 2025
      If the BIO RAD Power Supply shows error E 5, which means a power failure occurred during a timed run, and Power Failure detect (PFd) was activated, you can clear the error code by pressing any key (other than the key). The run was completed after the power failure.
  • M
    mike08Sep 5, 2025
    Why does BIO RAD PowerPac 200 Power Supply show E 6 Power Failure During a Timed Run error?
    • J
      Jessica ChoiSep 5, 2025
      If the BIO RAD Power Supply displays error E 6, indicating a power failure during a timed run, and Power Failure detect (PFd) was not activated, it means the run was terminated either due to a power failure or because the previous run was stopped by turning off the power supply or unplugging it. To clear the error code, press any key (other than the key).
  • M
    Michael GordonSep 8, 2025
    What to do if BIO RAD PowerPac 200 Power Supply displays E 7 Power Failure During an Untimed Run?
    • J
      Joshua SweeneySep 8, 2025
      If your BIO RAD Power Supply displays error E 7, which means a power failure occurred during an untimed run, it means the run was terminated either due to a power failure or because the previous run was stopped by turning off the power supply or unplugging the unit. To clear the error code, press any key (other than the key).