The UV source contained within the unit generates a light beam that traverses the sample
chamber and is accessible in the lamp chamber. Under normal use the lamp beam is confined within
the instrument. The unit should not be operated with the sample chamber lid open or the lamp
housing lid removed. Prolonged exposure to the beam may cause permanent eye damage.
There are no bio-hazardous materials within the unit; however this unit could be used with
bio-hazardous samples. Before using the instrument the customer should have in place
decontamination procedures designed to protect laboratory workers from occupationally acquired
infections. The sample chamber cell holders are removable and may be decontaminated using the
appropriate disinfectant for the bio hazard in question, rinsed with distilled water and then allowed
to dry. The sample chamber and exterior may be wiped with a suitable disinfectant cleaning wipe.
Decontamination. Equipment returned for repair should include an appropriate
decontamination certificate
It is the responsibility of the customer to ensure that the user of the equipment is provided
with a safe working environment.
Any chemicals used in Analyses should be used, stored and disposed of in accordance with
manufacturer’s guidelines and local safety regulations
Toxic Fumes. Efficient laboratory ventilation must be provided when working with volatile
solvents or toxic substances
Waste disposal. Disposal of some solvents and chemicals may be classed as hazardous waste
and must be dealt with in accordance with local regulatory practice.
Personal protective equipment. This is not required to operate the unit but the samples
measured may require PPE. A local risk assessment should be carried out.
This equipment may be connected and controlled from a PC. To preserve the integrity of
the measuring equipment it is essential that the attached PC itself conforms to basic safety and EMC
standards and is set up in accordance with the manufacturers’ instructions. If in doubt consult the
information that came with your PC. In common with all computer operation the following safety
precautions are advised.
To reduce the chance of eye strain, set up the PC display with the correct viewing position,
free from glare and with appropriate brightness and contrast settings
To reduce the chance of cross contamination from biological samples, use appropriate
personnel protection measures and disinfectant wipes on keyboard and mouse.