Biochrom Libra S50, S60, S70 & S80
UV Visible Spectrophotometers
Health & Safety Manual

Questions and Answers

  • M
    madisonmitchellAug 4, 2025
    Why are absorbance values higher than expected on my Biochrom Measuring Instruments?
    • R
      Richard BellAug 4, 2025
      Elevated absorbance readings can stem from several factors: * Verify you're using the correct cell type. * Ensure cells are free from fingerprints and contamination. * Confirm correct cell orientation. * Verify the reference used.
  • P
    peter99Aug 10, 2025
    What to do if absorbance value is lower than expected in biochrom Libra S50 Measuring Instruments?
    • S
      Sandra WrightAug 11, 2025
      If you observe lower than expected absorbance values, consider the following checks: * Ensure sufficient height of sample solution. * Verify that the sample and reference are not the same. * Make sure the sample compartment lid is properly closed.
  • K
    Kaitlyn SuttonAug 16, 2025
    How to fix negative absorbance reading on biochrom Libra S50 Measuring Instruments?
    • J
      James HallAug 16, 2025
      To resolve a negative absorbance reading, check that the sample and reference cells have not been swapped. Also, ensure there is sufficient height of sample solution. The beam height is nominally 15 mm above the cell base for normal measurements, and it is recommended that the solution be filled to 20 mm above the base.
  • B
    bobby21Aug 22, 2025
    How to improve poor reproducibility in nucleic acid analysis using Biochrom Measuring Instruments?
    • E
      erin49Aug 23, 2025
      To improve reproducibility in nucleic acid analysis, check for any particles in the solution and consider using background correction. Also, try increasing the concentration of the sample.
  • P
    Patrick MitchellAug 29, 2025
    What to check if Biochrom Measuring Instruments will not reference?
    • C
      Cheryl GrantAug 29, 2025
      If the instrument won't reference, check the orientation of the cuvette and verify that you are using the correct cell type.

Summary

Declaration of Conformity

Health and Safety

General Safety

Overview of safety directives and standards applicable to the equipment.

General Hazards

Identifies warning labels, intended users, and potential hazards of the instrument.

Unpacking and Installation

Instrument Connections

Equipment Operation

Controls and Indicators

Describes the function of the instrument's LEDs and indicators.

Intended Users

Specifies the qualifications and skills required for operating the instrument.

Instrument Preparation

Steps for preparing the instrument before sample measurement.

Post Run Procedures

Steps to follow after completing a sample measurement.

Performance Validation

Procedures for checking and ensuring the instrument's optical performance.

User Maintenance

Guidance on maintenance tasks that users can perform.

Troubleshooting

Customer Support and Service

Customer Support Contacts

Provides contact information for technical assistance and support.

Service, Repair or Return

Details the process for instrument servicing, repair, or return.

Disposal

Information and guidelines for the proper disposal of the instrument.

biochrom Libra S50 Specifications

General IconGeneral
Brandbiochrom
ModelLibra S50
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals