- 29 -
Door Sensor Calibration
With this menu, you calibrate the door sensor. On the following page, the calibration process is
explained step by step.
The measurement is an internal value for the distance between the magnet and the sensor.
When the door is door open, the voltage is approx. 2.5 V.
When the door is closed, the voltage moves towards 5 or 0 V depending on the direction of the
magnet.
Start calibration of door
sensor
INFO: During the calibra-
tion process, the controller
finds the set point automa-
tically on the basis of the
measurements taken with
door closed and door open.
INFO: The measurement
taken with door closed
should be between 10 and
100. The difference be-
tween door closed and door
open should be at least 10.