FLUOstar / POLARstar / LUMIstar OPTIMA Operating Manual BMG LABTECH
6/23 0413B0001I 2007-11-19
Time-resolved fluorescence: Limit of detection 70 amol / well europium
Spectral range (ex. and em.): 240 ... 740 nm
Dynamic range: 6 decades
Luminescence: Limit of detection <50 amol / well ATP
Spectral range 240 ... 740 nm
Dynamic range: 9 decades
Fluorescence polarization: Limit of detection: <5 mP SD at 1 nM fluorescein
Spectral range: (ex. and em.): 380 ... 740 nm
Dynamic range: 4 decades
Absorbance: Spectral range 240 ... 740 nm
Dynamic range 0.000-3.000 OD
Reproducibility 0.010 OD for 0-2 OD range,
0.030 OD for 2-3 OD range
Computer interface: RS232, 38400 baud, binary communication protocol
Power requirements: 100/115/230 V, 50/60 Hz
Consumption: max. 205 VA
Fuses: - T2.5A/250V for main power 230 V
- T3.15A/250V for main power 115 V
- T3.15A/250V for main power 100 V
(use original spare fuses provided by BMG LABTECH only)
Dimensions: Height: 27 cm, width: 44 cm and length 48 cm
Weight: 28 kg
Ambient conditions: Operating temperature: 15°C to 35°C
Storage temperature: -10°C to 50°C
Humidity of atmosphere: 20% to 80%
Non-condensing
Instrument conformity: Over voltage category II; contamination class II; protection class I
Robotic capabilities: Stacker for 50 microplates (optional)
Specifications are subject to change without notice.