EasyManua.ls Logo

BodyTrace BT005V - Error Codes

BodyTrace BT005V
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error Codes
Code Description Resolution
Er1 Battery low Replace batteries with four Duracell
AAs.
Er2 Movement
detected
Avoid moving or talking while
taking a measurement.
Er3 Overload The device is only accurate up to its
rated capacity.
Er4 No cellular
coverage
Move to another area, closer to a
window.
Use the device at a location where
you get strong cellular signal with
your mobile phone.
Er5 Cellular
connection error
Er6 Data
transmission
error
Er7 Internal error Replace batteries with four Duracell
AAs.
If the issue persists, contact
customer support.
Er9 Provisioning
error
Retake measurement.
If the issue persists, contact
customer support.
9