ED01_REV01-2020 IFU OB2012_2P-ENG
MEDICAL SUCTION UNIT
OPERATING INSTRUCTIONS

Table of Contents

Other manuals for Boscarol OB 2012 FA

Questions and Answers

  • D
    Dawn MasonAug 15, 2025
    What to do if vacuum power is low on Boscarol Medical Equipment?
    • K
      Kathryn AndersonAug 15, 2025
      If the vacuum regulator is completely open, close it entirely and check the vacuum on both the instrument and patient sides by turning the knob clockwise. If the filter is plugged, replace it. Ensure the tubing connecting the filter with the device is properly connected, not bent, and not plugged; replace if necessary and remove any bends. For an OB-J FA jar, disconnect the tubing, empty the jar, and verify the overflow valve moves correctly with the silicon gasket facing upwards, ensuring the jar remains upright. If the pump is damaged, contact an authorized service center.
  • D
    Donna WyattAug 18, 2025
    What to do if Boscarol OB 2012 FA Medical Equipment suction unit does not charge?
    • M
      Megan JohnsonAug 18, 2025
      If the suction unit does not charge when connected to the mains power supply, it could be due to a power supply failure. In this case, replace the adapter. Alternatively, contact an authorized service centre.
  • C
    Christina KellyAug 22, 2025
    How to fix Boscarol OB 2012 FA Medical Equipment suction unit that does not work?
    • S
      Stephanie ThomasAug 22, 2025
      If your Boscarol suction unit isn't working, the battery might be low. Try charging it with the charging cable or power supply adapter. If that doesn't work, the battery may have failed, or there could be an internal electronic circuit failure, and you should contact an authorized service center for assistance.
  • R
    rjonesAug 25, 2025
    Why Boscarol Medical Equipment works only when connected to power cable?
    • C
      Charles ChangAug 26, 2025
      If the suction unit works only when connected to the mains or external power cable, the fuses may have blown, so replace the fuse. Alternatively, the battery may be faulty, in which case you should contact an authorized service centre.
  • A
    abigail89Aug 28, 2025
    What to do if Boscarol Medical Equipment doesn't work on the wall bracket?
    • K
      Kristen HopkinsAug 28, 2025
      If the suction unit doesn't work when mounted on the wall bracket or with the external charger cable, the cable or wall bracket may be damaged, so replace the cable or wall-bracket. Check that the power supply voltage is within the required range (12-15 Vdc) and that the current is sufficient (12-15 Vdc - min. 6A). If those aren't the problem, there may be a suction unit internal circuit failure, and you should contact an authorized service centre.
  • C
    Craig BraunSep 1, 2025
    Why are battery lights off on my Boscarol Medical Equipment?
    • K
      Kevin PhillipsSep 1, 2025
      If the suction unit works, but the battery power indicator lights are off, it could be due to LED display or internal electronic circuit failure, so check that the LED display work if connected to the wall-bracket or to the external charger cable. If works, immediately charge the battery for at least 24 hours. If does not works, contact an authorized service centre. If the battery power is very low, charge the battery for at least 24 hours.
  • M
    mcdanielronaldSep 4, 2025
    Why is Boscarol OB 2012 FA battery life reduced?
    • B
      brendaharperSep 4, 2025
      If you notice a significantly reduced battery life in your Boscarol suction unit, it might be due to a faulty battery or an internal recharging circuit failure. Test the suction unit as described in page 12 of this manual or contact an authorised service centre.
  • A
    alison20Sep 7, 2025
    Why is vacuum always at maximum on Boscarol OB 2012 FA?
    • R
      Rebecca WilsonSep 8, 2025
      If the vacuum is always at maximum, even when the jar is removed, it indicates a fault on the internal pneumatic circuit. Contact an authorised service centre.
  • H
    hhendersonSep 12, 2025
    Why Boscarol OB 2012 FA has high noise and low suction?
    • J
      jacqueline78Sep 12, 2025
      If you experience high noise, low suction, and high vibration with your Boscarol device, it suggests internal pump damage. Contact an authorised service centre.

Summary

1. Intended Use

2. Warnings, Precautions and Important Information

3. Information Important to Know Before Use

4. Contraindications

6. Medical Suction Units OB 2012

6.4. Periodic Testing

Outlines the procedures for daily and six-monthly/yearly testing of the suction unit.

6.4.1. Daily Testing Procedure

Describes the steps for performing a daily functionality check of the suction unit.

6.4.2. Six-Monthly;Yearly Test

Details the comprehensive checks required for six-monthly/yearly device compliance.

6.5. Periodic Safety Maintenance

Specifies the schedule and importance of periodic safety maintenance for the unit.

6.6. Safety Information

Provides essential safety guidelines for users, patients, and third parties.

7. Jars for OB2012

8. Reuse, Cleaning and Disinfection

8.1. Reuse of Jar OB-J FA

Provides step-by-step instructions for dismantling and re-assembling the reusable OB-J FA jar.

8.2. Cleaning, Disinfection;Sterilization of Jar and Tube

Outlines procedures for cleaning, disinfecting, and sterilizing reusable jars and silicone tubes.

8.5. Cleaning Jar with SERRES Liners

Details the process for cleaning and managing the jar used with SERRES® disposable liners.

8.6. Disinfection;Sterilization of Jar and Tube

Provides instructions for disinfecting and sterilizing reusable jars and silicone tubes.

9. Accessories and Optional Parts

10. Internal Battery for Suction Units

13. Accessories, Consumables and Spare Parts

14. Technical Service

14.1. Troubleshooting

Addresses common malfunctions, their causes, and provides solutions for the suction unit.

15. Technical and Compliance Data

16. Electromagnetic Compatibility (EMC)

18. Warranty