Swiss LithoClast® Select
Swiss LithoClast
®
Select
MODE D’EMPLOI
DIRECTIONS FOR USE
Swiss LithoClast® Select System
FB-411_US_ed_2013_10_Not_Swiss LithoClast Select BSC.indd 1 6/3/2015 2:33:45 PM

Table of Contents

Questions and Answers

  • M
    Megan NielsenAug 12, 2025
    What to do if Medical Equipment by Boston Scientific shows Er 40 error?
    • S
      sanchezmeghanAug 12, 2025
      If the Medical Equipment by Boston Scientific displays error code Er 40, it indicates that the internal working pressure is not within the specified range, potentially due to input pressure being too high or too low. Here’s what you should do: 1. Verify and adjust the air pressure according to the value specified on the identification plate on the unit's back panel. 2. Reduce working pressure using the pneumatic rotary knob and try again by pressing the foot pedal. 3. Do not increase the pressure beyond 100%. If the error persists after these steps, further assistance is needed.
  • S
    stephaniewatsonAug 17, 2025
    How to fix Er 20 error on Medical Equipment by Boston Scientific?
    • S
      Shawn BarnesAug 17, 2025
      If the Medical Equipment by Boston Scientific displays error code Er 20, it means the internal ultrasound power value cannot be obtained, possibly due to a defective ultrasonic probe or handpiece. To resolve this, follow these steps: 1. Ensure the ultrasound probe is tightened. 2. Press the foot pedal to try again. 3. If the error reappears, replace the probe. 4. Press the foot pedal again. 5. If the error continues, reduce the ultrasound power to 80%. 6. Press the foot pedal again. 7. If the error persists, replace the ultrasound handpiece. 8. Press the foot pedal again. If the error is still present after these steps, further assistance is needed.
  • E
    Erin LewisAug 23, 2025
    What to do if Medical Equipment by Boston Scientific shows Er 30 error?
    • A
      Anthony AndrewsAug 23, 2025
      If the Medical Equipment by Boston Scientific displays error code Er 30, it indicates an internal power supply failure. Try turning the device off and then on again. If the failure continues, further assistance is needed.
  • A
    Angela SmithAug 30, 2025
    Why does my Medical Equipment by Boston Scientific show Er 10 error?
    • D
      Darlene PerezAug 30, 2025
      If the Medical Equipment by Boston Scientific displays error code Er 10, it may be due to a defective foot pedal or the foot pedal being accidentally pressed during startup. Ensure that you are not pressing the foot pedal when switching on the unit. Check for any obstructions that may impair the normal functioning of the pedal. If the issue persists, further assistance is needed.
  • T
    tylerramirezSep 4, 2025
    What causes Er 11 error on Medical Equipment by Boston Scientific?
    • T
      Tammy SanchezSep 4, 2025
      If the Medical Equipment by Boston Scientific displays error code Er 11, it could be due to a defective push button or the push button being accidentally pressed while switching on the unit. Make sure not to press any push buttons when turning on the unit. If the error continues, further assistance is needed.

Boston Scientific Swiss LithoClast Select System Specifications

General IconGeneral
Energy SourcePneumatic
Power Supply100-240 V AC, 50/60 Hz
CategoryMedical Equipment
CompatibilityCompatible with various endoscopes and probes
ApplicationUrology, Lithotripsy
Featuresuser-friendly interface
WeightApproximately 15 kg
Pressure Range1-4 bar
HandpieceErgonomic design
TypePneumatic
PortabilityPortable
FootswitchIncluded for hands-free operation
Probe OptionsMultiple sizes available