Patient Screening and Preparation
ImageReady™ MRI Full Body Guidelines for WaveWriter Alpha™ and WaveWriter Alpha™ Prime Spinal Cord Stimulator Systems
92395577-02 13 of 797
en
Patient Screening and Preparation
The following table summarizes the WaveWriter Alpha Systems/Patient-related Conditions of Use that
must be met in order for an MR Conditional scan to be performed. For each condition or requirement,
suggested methods to determine eligibility are listed. It is not required to use all suggested methods. Any or
a combination of the suggested methods may be used.
Appendix A, “ImageReady MRI Full Body Patient Eligibility,” contains a form that may be used by the
physician managing the patient’s SCS system to conrm the patient meets the SCS Implant System
Conditions for MRI Scans as described in this manual.
Table 2. WaveWriter Alpha Systems/Patient Screening and Preparation Conditions
# Condition for Scanning Suggested Methods to Determine
Eligibility
1. The patient is implanted with a WaveWriter Alpha SCS
System composed only of components listed in “Table 1.
Components that are eligible for WaveWriter Alpha System
with ImageReady MRI Full Body Technology” on page 3 of
this manual.
Note: Leads should be connected directly into the IPG.
Patient should not be implanted with Lead Extensions,
Splitters, or Adapters.
Check patient records
Check the Patient ID card
Check model numbers in Table 1 of this manual or
by contacting Boston Scientic Neuromodulation
Technical Services.
Conrm with the physician responsible for
managing the Patient’s SCS System.
2. The Lead implant location is epidural. Check patient records
Verify by X-Ray
3. The patient has no abandoned Leads or IPGs (Leads
or IPGs that are not connected to the functioning WaveWriter
Alpha System).
Check patient records
Verify by X-Ray
4. The IPG is implanted in the upper buttock or the lower ank Check patient records
Examine the patient by palpation to determine the
location of the IPG
Verify by X-Ray
5. For rechargeable systems, IPG is fully charged prior to the
MRI scan.
Make sure three bars are displayed
at the
top right of the Home screen on the Remote Control.

Table of Contents

Other manuals for Boston Scientific WaveWriter Alpha Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Boston Scientific WaveWriter Alpha Series and is the answer not in the manual?

Boston Scientific WaveWriter Alpha Series Specifications

General IconGeneral
BrandBoston Scientific
ModelWaveWriter Alpha Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals