Questions and Answers

  • T
    Timothy GregoryAug 1, 2025
    What to do if Bovie Medical Equipment shows error F2?
    • J
      Joseph SalinasAug 2, 2025
      If the Bovie Medical Equipment displays error code F2, release either the cut or coag button on the handpiece, or the cut or coag pedal on the footswitch.
  • J
    jamescraigAug 6, 2025
    What to do if Bovie Aaron 1250 shows error E7?
    • D
      Dr. Scott ToddAug 6, 2025
      If the Bovie Medical Equipment displays error code E7: 1. Turn the unit off. 2. Allow the unit to cool for 20 minutes. 3. Turn the unit on. 4. If the error code reappears, record the number and contact Bovie® customer service.
  • D
    deckeranthonyAug 11, 2025
    What to do if Bovie Aaron 1250 shows error E6?
    • E
      Erin HicksAug 11, 2025
      If the Bovie Medical Equipment displays error code E6: 1. Turn the unit off. 2. Verify that the unit is connected to the line voltage. 3. If the error code reappears, record the number and contact Bovie® customer service.
  • P
    petersenjasonAug 16, 2025
    What to do if Bovie Medical Equipment shows error F1?
    • P
      Patty NortonAug 16, 2025
      If the Bovie Medical Equipment displays error code F1: 1. Turn off, then turn on the generator. Do not press buttons or activate accessory devices during the self-test. 2. If the error code reappears, disconnect all accessories. Turn off, then turn on the generator again. 3. If the problem persists, replace the handpiece or footswitch and repeat the restart. 4. If the error code reappears, record the number and call Bovie® customer service.
  • K
    Kimberly Hull MDAug 21, 2025
    What to do if Bovie Aaron 1250 shows error E4?
    • T
      Tonya JacobsonAug 21, 2025
      If the Bovie Medical Equipment displays error code E4: 1. Turn the unit off. 2. Turn the unit on. 3. If the error code reappears, record the number and contact Bovie® customer service.
  • M
    Morgan HerreraAug 27, 2025
    Why won't the footswitch activate my Bovie Medical Equipment?
    • J
      Jonathan HarrisAug 27, 2025
      The Bovie Medical Equipment will not allow the footswitch to activate the unit if Bipolar footcontrol is selected, but the Bipolar Mode is not selected.

Summary

Introducing the Aaron 1250 Electrosurgical Generator

Safety and Contraindications

Provides critical safety warnings, hazards, and lists any conditions where use is not advised.

Controls, Indicators, and Receptacles

Getting Started

Initial Setup and Function Checks

Covers initial inspection, installation, and functional verification of the generator and its modes.

Using the Aaron 1250

Setup Safety and Procedure Preparation

Covers pre-operation safety, setting up the unit, and preparing for monopolar or bipolar surgery.

Activation Safety and Operation

Details critical safety measures during activation and step-by-step guides for unit activation.

Maintaining the Aaron 1250

Troubleshooting

Error Code Resolution

Lists common error codes, their descriptions, and recommended actions for resolving issues.

Technical Specifications

Bovie Aaron 1250 Specifications

General IconGeneral
TypeElectrosurgical Generator
Output Power120 watts
Line Voltage100-240 VAC
ModesCut, Coagulation, Bipolar
Cut ModeBlend
Coagulation ModeDesiccation, Fulguration