Table of Contents

Questions and Answers

  • E
    Eric MartinezAug 2, 2025
    What to do if the Bovie Medical Equipment shows a line voltage error?
    • D
      David SmithAug 2, 2025
      If you see a line voltage error on your Bovie Medical Equipment, follow these steps: First, turn the unit off. Second, check that the unit is connected to the correct line voltage. Finally, if the error code still appears, record the error number and contact Bovie Medical Corporation customer service.
  • A
    Andrew MillerAug 12, 2025
    What to do if the Bovie Medical Equipment overheats?
    • M
      Michelle WigginsAug 13, 2025
      If your Bovie Medical Equipment indicates that the internal temperature has exceeded its limit, here's what to do: First, turn the unit off. Second, let the unit cool down for 20 minutes. Third, turn the unit back on. If the error code reappears after this, record the number and then contact Bovie Medical Corporation customer service.
  • M
    Melissa TurnerAug 22, 2025
    What to do if handswitch or footswitch pedal is stuck on Bovie Medical Equipment?
    • T
      Terry GomezAug 23, 2025
      If the handswitch or monopolar footswitch pedal on your Bovie Medical Equipment seems to be stuck, try the following: 1. Turn off, then turn on the generator, ensuring no buttons are pressed or accessory devices activated during the self-test. 2. If the error code returns, disconnect all accessories, then turn the generator off and on again. 3. If the problem continues, replace the handpiece or footswitch and repeat the restart process. 4. Should the error code still reappear, record the number and contact Bovie Medical Corporation customer service.

Summary

Introducing the Aaron 950 Electrosurgical Generator

Safety Warnings

Crucial warnings for the safe and effective use of electrosurgical equipment.

Fire and Electric Shock Hazards

Details significant hazards related to fire, explosion, and electrical shock during operation.

Patient Safety and Cautions

Provides essential cautions and notices for patient safety and proper equipment handling.

Controls, Indicators, and Receptacles

Cut, Blend, and Coag Controls

Explains controls for Cut, Blend, and Coagulation modes, including power settings.

Fulguration, Bipolar, and Preset Controls

Details controls for Fulguration, Bipolar modes, and user-defined presets on the front panel.

Getting Started

Using the Aaron 950

Setup Safety and Procedures

Crucial safety measures and step-by-step guide for setting up the generator for operation.

Activation Safety Precautions

Precautions to prevent hazards during unit activation and operation.

Maintaining the Aaron 950

Troubleshooting

Error Code Resolution

Lists common error codes, their descriptions, and recommended actions for resolving issues.

Repair Policy and Procedures

Technical Specifications

Warranty Information

Bovie Aaron 950 Specifications

General IconGeneral
TypeElectrosurgical Generator
Frequency500 kHz
Line Frequency50/60 Hz
ModesCut, Blend, Coagulation
Line Voltage100-240 VAC
DisplayLED