OUTPUT CHARACTERISTICS
Maximum Output for Bipolar and Monopolar Modes
Power readouts agree with actual power into rated load to within 20% or 5 watts, whichever is greater.
A-8 Bovie Medical Corporation
Mode Output Power Output Frequency Repetition Rate Vpeak
Cut 60 W @ 500 Ω 357 kHz ± 50 kHz N / A 1.5 KV
Blend 60 W @ 800 Ω 357 kHz ± 50 kHz 30 kHz ± 5 kHz 2.0 KV
Coagulation 60 W @ 1000 Ω 375 kHz ± 50 kHz 60 kHz ± 5 kHz 3.8 KV
Fulguration 35 W @ 1000 Ω 575 kHz ± 50 kHz 30 kHz ± 5 kHz 6.5 KV
Bipolar 30 W @ 200 Ω 520 kHz - 14kHz + 50 kHz 19 kHz ± 5 kHz 2.0 KV