STANDARDS AND IEC CLASSIFICATIONS
Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation failure because of the way in which
they are connected to the protective earth conductor.
Type BF Equipment (IEC 60601-1) / Defibrillator Proof
The Aaron 950™ Electrosurgical Generator provides a high degree of protection against electric shock,
particularly regarding allowable leakage currents. It is type BF equipment. Patient connections are isolated
from earth and resist the effects of defibrillator discharge.
Drip Proof (IEC 60601-2-2)
The generator enclosure is constructed so that liquid spillage in normal use does not wet electrical insulation or other
components which, when wet, are likely to affect adversely the safety of the generator.
Electromagnetic Interference
When other equipment is placed on or beneath an activated Bovie Medical Corporation electrosurgical generator, the
Aaron 950™ Electrosurgical Generator operates without interference. The generator minimizes electromagnetic inter-
ference to video equipment used in the operating room.
Electromagnetic Compatibility (IEC 60601-1-2 and IEC 60601-2-2)
The Aaron 950™ Electrosurgical Generator complies with the appropriate IEC 60601-1-2 and IEC 60601-2-2
specifications regarding electromagnetic compatibility.
Voltage Transients (Emergency Generator Mains Transfer)
The Aaron 950™ Electrosurgical Generator operates in a safe manner when the transfer is made between line AC
and an emergency generator voltage source.
EMC COMPLIANCE
Special precautions should be taken regarding the Aaron 950™. Medical Electrical Equipment needs special precau-
tions regarding EMC and needs to be installed and put into service according to the EMC information provided in
this manual.
Understand that only the Accessories supplied with or ordered from Bovie® should be used with your device. The
use of accessories, transducers, and cables other than those specified, may result in increased Emissions or decreased
Immunity of the Aaron 950™. The Aaron 950™and its accessories are not suitable for interconnection with other
equipment.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment. The Aaron 950™
should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the
Aaron 950™ should be observed to verify normal operation in the configuration in which it will be used.
A-4 Bovie Medical Corporation