A-5User’s Guide • Aaron 950™
Recommended separation distances between portable and mobile RF
communications equipment and the Aaron 950™
The Aaron 950™ is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Aaron 950™ can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Aaron 950™ as recommended below, accord-
ing to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =
[
3.5
]
√
P
V
1
80 MHz to 800 MHz
d =
[
3.5
]
√
P
E
1
800 MHz to 2.5 GHz
d =
[
7
]
√
P
E
1
0.12 0.12 0.23
0.38 0.38 0.73
1.2 1.2 2.3
3.8 3.8 7.3
12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Guidance and manufacturer’ s declaration – electromagnetic emissions
The Aaron 950™ is intended for use in the electromagnetic environment listed below. The customer
or the user of the Aaron 950™ should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF Emissions CISPR 11 Group 2
The Aaron 950™ must emit
electromagnetic energy in order
to perform its intended function.
Nearby electronic equipment
may be affected.
RF Emissions CISPR 11 Class A
The Aaron 950™ is suitable for
use in all establishments other
than domestic and those directly
connected to the public
low-voltage power supply net-
work that supplies buildings used
in domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies