PERFORMANCE CHARACTERISTICS
Input Power
Duty Cycle
Under maximum power settings and rated load conditions (Pure Cut, 60 watt @ 500 load), the generator is
suitable for activation times of 10 seconds on, 30 seconds off for one hour.
Dimensions and Weight
Operating Parameters
Transport and Storage
A-2 Bovie Medical Corporation
120 VAC ± 10% 220 VAC ± 10%
Mains line frequency: 50 / 60 Hz Mains line frequency: 50 / 60 Hz
Power consumption: 240 VA Power consumption: 240 VA
Fuses (two): 2.0 A (Slow Blow) Fuses (two): 1.0 A (Slow Blow)
Ambient temperature range -40° to +70° C
Relative humidity 10% to 100%, including condensation
Atmospheric pressure 50kPa to 106kPa
Width 26 cm (10.25 in.) Depth 11.4 cm (4.5 in.)
Height 17.8 cm (7.0 in.) Weight < 3.2 kg (< 7 lbs)
Ambient temperature range +10° to +40° C
Relative humidity 30% to 75%, non-condensing
Atmospheric pressure 70kPa to 106kPa
Warm-up time
If transported or stored at temperatures outside the operating temperature range, allow one hour
for the generator to reach room temperature before use.